BARNSTEAD|THERMOLYNE
NANOpure
®
DIamond
UV
ultrapure water system
Operation Manual
Series 1191
Model No. Voltage
D11911 100V - 240V
LT1191X1 • 6/22/00
Vista de página 0
1 2 3 4 5 6 ... 55 56

Resumo do Conteúdo

Página 1 - NANOpure

BARNSTEAD|THERMOLYNENANOpure®DIamond™UVultrapure water systemOperation ManualSeries 1191Model No. VoltageD11911 100V - 240VLT1191X1 • 6/22/00

Página 2 - Table of Contents

Other AccessoriesOptional accessories for this unit include a float switch,low pressure switch, N.I.S.T. Traceable CalibrationModule, remote dispenser

Página 3 - Safety Information

UV Lamp InstallationLocate the UV chamber inside the left door of theNANOpure DIamond UV. Install the UV Lamp as follows:1. Disconnect the power cord

Página 4 - SAFETY INFORMATION

Bench Mounting1. Place the NANOpure DIamond UV on a benchtop that is accessible to pretreated water, elec-tricity and an atmospherically vented drain.

Página 5 - Introduction

Installing the Control Panel in aRemote LocationFor your convenience, the control panel can be removedfrom the unit and mounted at a convenient locati

Página 6 - General Specifications

14Water ConnectionsFeed Water Connection1. Locate the length of 3/8” (.95 cm) O.D. tubingprovided with the unit with a quick disconnectinsert on one e

Página 7 - Declaration of Conformity

15Main Power SwitchThe main power switch on the NANOpure DIamond UV islocated on the lower back right side of the unit (as youface the front of the un

Página 8 - Unpacking

SwitchesWhen the main power switch (on the lower back right sideof the unit as you face the unit) is ON, the switches onthe control panel function as

Página 9 - Installation

17Cartridge packs will come bagged with four manifold connection caps.Part ApplicationDIamond Kit Organic Free R/O & Distilled FeedD50280 Ultra-Lo

Página 10 - Other Accessories

18Initial Rinse1. Attach the feed water line to the unit by snapping the quick disconnect coupling into the quick disconnect body in the lower left ba

Página 11 - UV Lamp Installation

19System Sanitization ProcedureYour NANOpure DIamond UV has been shipped with acleaning syringe (Cat. No. CMX25). When the system isnew or whenever a

Página 13 - Remote Location

20and press ENTER. Display will now read, “UnitSanitizing.”14. Unit may now be left unattended until displayreads, “Sanitization Complete: Press ENTER

Página 14 - Water Connections

211. Turn the system power ON by depressing themain power switch to the “I” position.2 The system greeting display indicates the typeof unit you own.

Página 15 - Controls

Adjusting Display Brightness1. From the (Idle) mode, when the display reads,“NANOpure Diamond (Idle) xx:xx:xxxx/xx/xx,” press the UP or DOWN arrow unt

Página 16 - Switches

232. Press ENTER.3. Press ENTER when the unit displays “PrinterAttached?”4. When prompted with “Date Format” press theUP or DOWN arrow to show the dat

Página 17 - Initial Operation

242. Unlatch and open the right side door.3. Find and copy down the two cell constantsattached to the gray cable, leaving the electron-ics board mount

Página 18 - Initial Rinse

25Use of Standby ModeAt the end of the work day, place the NANOpureDIamond UV in Standby mode for the night. To place theunit in Standby mode:1. From

Página 19 - System Sanitization Procedure

267. Press ENTER.8. Press BACK to return to the (Idle) mode.The second line of the display will show a “Below Setpoint” message once every 8 to 10 sec

Página 20 - INITIAL OPERATION

27USER SETTINGS3. Press the UP or DOWN arrow to select “Yes.”4. Press ENTER and the timer will be reset forapproximately five months of operational ti

Página 21 - Normal Operation

28NotePress the STOP/STOP switch to can-cel the flush.Setting Timed Dispensing1. From the normal recirculating mode, when the dis-play is showing puri

Página 22 - User Settings

292. Press the UP or DOWN arrow to select “MΩ-cm”or “uS/cm” Press ENTER.3. Press BACK to return to the normal recirculatingmode.Temperature Compensati

Página 23 - Setting the Cell Constant

Your Barnstead|Thermolyne NANOpure DIamond UVultrapure water system has been designed with function,reliability, and safety in mind. It is your respon

Página 24 - USER SETTINGS

30Accessories D8964 (float switch) and D2706 (pressureswitch) are designed to protect the NANOpure DIamondUV pump by alerting the system of an inadequ

Página 25 - Selecting the Set Point

31Performing an ElectronicCalibration Using the OptionalN.I.S.T. Calibration ModuleIf you purchased the optional N.I.S.T. calibration module(Catalog N

Página 26 - Resetting the UV Timer

32N.I.S.T. CALIBRATION MODULE11. Press BACK to return to the (Idle) mode.12. Turn the main power switch to the OFF “O” position. Unplug the unit.13. R

Página 27 - Setting Volumetric Dispensing

Connecting NANOpureDIamond to Computer andCommunicating Though the RS-232 Port UsingHyperterminal or ProcommHyperterminal RS-232 Capture Instructions1

Página 28 - Changing Purity Units

34capture text. Accepting default will put the filewhich you name in the c:/windows directorycalled capture.txt and the file will be an ASCIItext file

Página 29 - Temperature Compensation

3512. Type in 21 <Enter>, sets to COM2: or (Type in20 <Enter>, sets to COM1:), depending on yourcomputer.13. Type in 24 <Enter>, sav

Página 30 - Installing Float or Pressure

367. From the “Printer Attached” display, press theUp or DOWN arrow for (YES).8. Press ENTER. Press BACK twice to return tothe (Idle) mode.9. During n

Página 31 - N.I.S.T Calibration Module

37down; or—the paper may be jammed in the printer. Turnthe printer OFF before checking for jammed paper.Remove the paper jam and turn the printer back

Página 32 - Remote Dispenser

38General Cleaning InstructionsDisconnect electrical service to the unit. Wipe exteriorsurfaces with lightly dampened cloth containing mild soapsoluti

Página 33 - Computer/Printer Setup

System SanitizationAfter sanitization is complete (approximately 3.5 hours),place the unit into its normal (recirculation) mode to allowthe system to

Página 34 - COMPUTER/PRINTER SETUP

3. Avoid splashing disinfecting solutions on clothing or skin.4. Ensure all piping connections are tight to avoidchemical leakage.5. Ensure adequate v

Página 35 - Printer

4014. Unit may now be left unattended until the display reads, “Sanitization Complete: PressENTER.”15. After sanitization is complete and the userpres

Página 36 - Printer Setup

41Cartridge Pack ReplacementThe frequency with which you will need to clean your unitand replace your cartridge pack is dependent on yourfeed water’s

Página 37 - Installing the Paper Roll

0.2 Micron Filter ReplacementReplace the 0.2 micron final filter whenever any of the fol-lowing conditions occur: every 30 days, the product waterflow

Página 38 - Maintenance and Servicing

43UV Lamp ReplacementThe ultraviolet lamp requires changing every 5-6 months.Lamp Life will vary according to the number of times theNANOpure DIamond

Página 39 - System Sanitization

446. Make sure the UV lamp cable is tucked behindthe chamber before closing and relatching thedoor.7. Reconnect power cord and turn unit ON.8. Reset U

Página 40 - System Depressurization

45Problem Possible Causes SolutionsNANOpure DIamond completely No electrical power to NANOpure Ensure NANOpure DIamond power inactive (pump and displ

Página 41 - Cartridge Pack Replacement

46Poor quality feed water. If a Barnstead ROpure is the feed water source, check that the membrane is functioning properly.If a Barnstead Still is the

Página 42 - 0.2 Micron Filter Replacement

47Display reads, “Meter PC board communication error. Turn the NANOpure DIamond OFFBoard Problem.” (I/O Switch) and restart the unit.System electronic

Página 43 - UV Lamp Replacement

48Recommended SparesConsumablesConsumable parts are those required to support the day-to-day operation of this equipment.Barnstead|Thermolyne establis

Página 44 - Shutdown

49Safety StockFor critical applications where performance with minimum downtime is required, Barnstead|Thermolyne rec-ommends that you maintain a loca

Página 45 - Troubleshooting

Congratulations on your purchase of a Barnstead|Thermolyne NANOpure DIamond UV ultrapure water sys-tem. This water purification system is designed to

Página 46 - TROUBLESHOOTING

50Flow ChartUnit Initialization & MeterCalibration CheckNANOpure DIamond (Idle)xx :xx:xx xx /xx/xxstart_stopkey pressedup or downkeys pressedSani

Página 47

51Wiring DiagramCONNECTORREMOTE DISPLAYCSW1190X2SW1190X2SW1190X2(SPG)230V: (2) FSX54& FUSES:120V: (2) FSX47& D11931 ONLYMODEL #s D11911 LINENE

Página 48 - Replacement Parts

52Ordering ProceduresPlease refer to the Specification Plate located inside theleft side door for the complete model number, serial num-ber, and serie

Página 49 - Safety Stock

53Barnstead|Thermolyne Corporation warrants that if a product manufactured by Barnstead|Thermolyne andsold by it within the continental United States

Página 52 - Ordering Procedures

9001ISOREGISTEREDBarnstead|Thermolyne2555 Kerper Blvd.P.O. Box 797Dubuque, IA 52004-2241Phone: 319-556-2241 800-553-0039Fax: 319-589-0516E-Mail Addr

Página 53 - One Year Limited Warranty

Dimensions and Clearance RequirementsDimensions 13.4” W x 19.5” H x 16.9” D (34.0 x 49.5 x 42.8 cm)ClearancesSides - 9” (22.9 cm) minimum for servicin

Página 54

Environmental Conditions Operating: 4°C - 49°C; 20% to 80% relative humidity, non-condensing. Installation Category II (over-voltage) in accordance w

Página 55

1. Remove the unit from its shipping container.Remove all contents carefully. Ensure that theUV lamp, feed and drain tubing, sanitizationcartridge, w

Página 56 - Barnstead

9Choosing a SiteThe NANOpure DIamond UV features a removable con-trol panel display which allows the system to be mountedalmost anywhere within the la